dc.contributor.author | Belgin, Ezgi Eren | |
dc.contributor.author | Aycık, G. A. | |
dc.date.accessioned | 2020-11-20T15:06:36Z | |
dc.date.available | 2020-11-20T15:06:36Z | |
dc.date.issued | 2015 | |
dc.identifier.issn | 1350-4487 | |
dc.identifier.uri | https://doi.org/10.1016/j.radmeas.2014.10.003 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12809/3151 | |
dc.description | WOS: 000349725500006 | en_US |
dc.description.abstract | Simulation is a simple method that allows for facile estimation of a detector response function when the specifications of the investigated sample and the sample-detector geometry are known. In this study, the dependence of the full-energy peak efficiency on the sample and detector specifications was studied via analytical techniques and the Monte Carlo approach in the energy range of 150-2000 keV for a well-type measuring geometry, and a semi-empirical model was developed. The modeling of the detector efficiency is described in detail. The compute results were compared with experimental data, and comparison of the calculated efficiencies with the measured values indicated that the deviations between the calculated and experimental efficiencies were mostly less than +/- 10%. Therefore, the simulation model appear to be suitable for routine environmental radionuclide analysis when uncertainties of up to 10% are acceptable for energies between 150 and 2000 keV. (C) 2014 Elsevier Ltd. All rights reserved. | en_US |
dc.description.sponsorship | Mugla Sitki Kocman University Scientific Research Projects DepartmentMugla Sitki Kocman University [10/01] | en_US |
dc.description.sponsorship | We are especially grateful to Dr. O. Akpolat for his invaluable assistance with MATLAB programming. We are also grateful to Mugla Sitki Kocman University Scientific Research Projects Department for their sponsorship (Project no. 10/01). | en_US |
dc.item-language.iso | eng | en_US |
dc.publisher | Pergamon-Elsevier Science Ltd | en_US |
dc.item-rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Monte Carlo Simulation | en_US |
dc.subject | Efficiency Calibration | en_US |
dc.subject | Efficiency Modeling | en_US |
dc.subject | Hpge Detector | en_US |
dc.subject | Gamma Spectrometer | en_US |
dc.title | Derivation of an efficiency-calibration simulation for a well-type HPGe detector using the Monte Carlo approach and analytical techniques | en_US |
dc.item-type | article | en_US |
dc.contributor.department | MÜ, Fen Fakültesi, Kimya Bölümü | en_US |
dc.contributor.institutionauthor | Belgin, Ezgi Eren | |
dc.identifier.doi | 10.1016/j.radmeas.2014.10.003 | |
dc.identifier.volume | 73 | en_US |
dc.identifier.startpage | 36 | en_US |
dc.identifier.endpage | 45 | en_US |
dc.relation.journal | Radiation Measurements | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |